| Item |
Units |
Specifications |
Flow rate range (Full scale)
Ethanol equivalent |
g/min |
0.05, 0.1, 0.2, 0.5, 1.0
2.0, 5.0, 10 |
| Accuracy |
%F. S. |
±1 |
| Linearity |
%F. S. |
±0.5 |
| Repeatability |
%F. S. |
±0.2 |
| Pressure capacity |
kPa |
1000 |
| Operation temperature |
°C |
15-35 |
| Temperature coefficient |
%F. S. /°C |
Zero 0.1
Span 0.1
|
| Leak integrity |
Pa·m3/sec |
1x10-11 |
Materials exposed
to fluids |
Body
Seals
Flow sensor |
SUS-316L
Au
SUS-316L
|
| Standard fittings |
Inlet
Outlet
|
3.2mm VCR Male
3.2mm VCR Male
|
| Flow rate command |
Volts DC |
0-5 |
| Flow rate output |
Volts DC |
0-5 |
| Power supply requirements |
 |
+15V, 100mA
-15V, 200mA
|
| Alarm display |
Normal
Condition change
Abnormal |
Green LED flash
Red LED flash
Red LED continuous on |
| Zero adjustment |
 |
Zero switch or
auxiliary input |
| Digital communication |
Digital interface
Alarm output
Zero adj. Input
Preset input |
RS-232C
Open collector output
Available
Available
|
| Connector and cable |
Main connector
Communication
Cable (Optional) |
D05 8 pin
CC-DOSF (D-SUB 9 pin)
CC-T2 (D-SUB 25 pin) |
| Mounting direction |
 |
Free |
| Weight |
g |
800 |
| Optional |
LEP
CP
PT |
Surface polishing
Anti-corrosive plating
Pt Seal |